Iontophoresis Device Idromed4 - Problems and how to solve them

Impressum/Imprint      Sweat side DeutschSweating site francais

- Index - Order form - Iontophoresis forum -   Manual - how to use iontophoresis - sweating-betty

Hyperhidrosis or extreme Sweating isnt normal !

This is a sample of the content of the manual from the iontophoresis-device Idromed4 - Note: we cannot guarantee the actuality and accuracy.

If you buy a Idromed iontophoresis device a actual maual is included.

The manual is spreaded over several pages.

With the idromed® 4 you have purchased a tried and tested iontophoresis unit for the treatment of excessive sweating (hyperhidrosis) on hands, feet and armpits (with optional accessories). We hope that you will be satisfied with the use of the idromed® 4. In order to achieve good results, please follow these instructions exactly.

CONTENT

1.    Important information on the idromed® 4   
1.1.    Safety information 
2.    Scope of delivery   
3.    Application of the idromed® 4  
3.1.    Fields of application  
3.2.    Preparation of the treatment   
3.3.    Treatment
3.3.1.    Treatment of hyperhidrosis on the hands  
3.3.2.    Treatment of hyperhidrosis on the feet   
3.3.3.    Treatment of hyperhidrosis on hands and feet   
3.3.4.    Treatment of hyperhidrosis at the armpits   
4.    Important tips and information  
5.    Cleaning and Care   
6.    Maintenance   
7.    Problems and how to solve them   
8.    Technical Data  
9.    Warranty and liability 

7 Problems and how to solve them

Problem Solution
The lamp on the mains plug does not work. In brightly lit rooms, it can be difficult to see whether the lamp is on or not. Shade the lamp with your hand to check, whether it is working properly.

The socket may be defect. Please connect the unit to another socket or check whether the socket in question is working properly.

The green control lamp ( I ) does not work. In brightly lit rooms, it can be difficult to see whether the lamp is on or not. Shade the lamp with your hand to check, whether it is working properly.

The unit is not connected properly. Check the connection between the mains cable and the unit.

The unit is not switched on. Turn the control knob to the value 1 and switch on the unit via the foot switch.

The orange error lamp is on. In case of a functional error, the output current of the unit will automatically be cut off and the orange lamp will give light. Switch off the unit and switch it on again. If the error lamp is still alight, please contact the Customer Service.
The LCD-display shows 0 mA. The LCD-display shows a positive value only when the electric circuit is closed.

If no current flow is indicated during the treatment, check the connection between the mains cable and the unit as well as the electrode connections.

If there isn’t a positive value, if all connections are correct, please contact the Customer Service.

Current intensity is not sensible. Check the connections and the function of the unit (see above).

 

If you have further problems, please contact our Service-Hotline.